A 77-year-old man with severe aortic stenosis developed life-threatening complications 12 days after an apparently successful transcatheter aortic valve replacement, requiring a second valve implanted inside the first to stop massive leakage and hemolysis.
The patient received a 26-mm Edwards SAPIEN 3 Ultra RESILIA valve via transfemoral access. Completion angiography showed no paravalvular leak (PVL) or central aortic insufficiency. Transthoracic echocardiography on postoperative day 1 confirmed trace PVL, a mean gradient of 11 mm Hg, and preserved left ventricular ejection fraction at 60%, according to the case report published in Cureus.
Twelve days after discharge, he returned with progressive dyspnea, orthopnea, dark urine, and acute anemia — hemoglobin 6.8 g/dL, down from 9.2 g/dL at discharge. Labs revealed lactate dehydrogenase 1,695 U/L, total bilirubin 1.4 mg/dL (predominantly unconjugated), and haptoglobin below detection, consistent with intravascular hemolysis.
Transesophageal echocardiography demonstrated severe eccentric PVL, left ventricular dilation with reduced systolic function, and moderate-to-severe mitral regurgitation. The prosthetic valve had migrated toward the left ventricle. Aortic root angiography confirmed a lower position than at implant.
A heart team elected for rescue transcatheter valve-in-valve (TAV-in-TAV) implantation with a second 26-mm Edwards SAPIEN 3 Ultra RESILIA valve. Post-procedure imaging showed no significant PVL.
Mitral regurgitation improved from moderate-to-severe to mild. Hemolysis resolved rapidly with no further transfusions. The patient was extubated on postprocedural day 1 and the intra-aortic balloon pump was removed on day 2.
At nine-month follow-up, left ventricular ejection fraction had recovered to 55%, with mild residual PVL and mitral regurgitation.
“Delayed migration of an initially well-positioned transcatheter valve is an uncommon complication that can result in new loss of prosthesis-annulus apposition and clinically significant PVL,” the authors wrote in Cureus.
They said that early migration presenting with severe PVL, hemolytic anemia, and secondary mitral regurgitation is less well described than late migration causing stenosis or regurgitation years after implant.
The report highlights the diagnostic value of transesophageal echocardiography when transthoracic findings do not explain clinical deterioration, and the potential for recovery of ventricular function and functional mitral regurgitation after correction of the mechanical lesion.
The case was previously presented as a poster at the Society of Cardiovascular Anesthesiologists’ 48th Annual Meeting in Nashville, April 23–26, 2026.